Add the Correction Weight and Take a Residual Measurement

The residual measurements are taken with the correction weight or the correction
weight and trim weights attached to the machine. The correction weight should
cancel out the initial unbalance. A residual vibration measurement is taken to
measure the remaining unbalance. Trim weights are added to the machine to
cancel out the vibration measured during a residual measurement.

The Dynamix 2500 data collector calculates and displays a recommended
balancing weight in the Correction Weight - Solution screen.

You can split the correction weight into two distinct angles, as well as modify the
radius of the correction weight. Additionally, the Dynamix 2500 data collector
calculates the amount of weight that might be removed if weights are already
attached to the rotor.

When you need to save the Correction Weight solution, press F3 (Memory.

F3 (Memory) take you to
the Save Setup screen if
you want to save your
Solution.