Rockwell Automation 1441-DYN25-MBAL Balancing Extension Module for the Dynamix 2500 User Manual

Page 49

Advertising
background image

Rockwell Automation Publication 1441-UM004A-EN-P - May 2011

49

Balancing Extension Module Chapter 2

3.

In the Sensitivity field, enter the transducers sensitivity in millivolts (mv)
per Engineering Unit (EU).

100 mV/EU is used for most acceleration transducers, 200 mV/EU for
most non-contact displacement transducers, and 1000 mV/EU if the input
is volts and the scales is to be read directly. The default setting is 100 mV/
EU.

4.

In the Input Range field, select the signal input range.

The range is between Automatic and a fixed range in transducer units in a
1 - 2 - 5 sequence. The available fixed values vary depending on the
transducer type being used.

5.

In the Type field, select the appropriate weight position type.
Continuous specifies that weights may be placed at any placement

position on the balance plane.

Fixed (Works with Number of Positions and Position 1 Offset fields.)

specifies that weights may be placed only at designated positions, for
example, rotors having a specific weight hole pattern, or with a specific
number of fan blades.

6.

In the No. of posns field, enter the number of weight positions for the
respective plane, from 3…360.

Applies to Fixed weight position measurements only. The first position is
assumed to be at zero degrees, subsequent positions are evenly spaced.

7.

In the Position 1 Offset field, enter the fixed position 1 offset for the
respective plane, from 0…360.

Applies to Fixed weight position measurements only.

8.

Press F3 (Start) or Enter to begin the two-plane balancing procedure.