How to Perform Two-plane Balancing

After setting up your balance measuring equipment, and setting the data
collector’s balancing measurement parameters, you are ready to perform the
two-plane balancing procedure.

Perform the Initial Runs

The two plane procedure is slightly different, depending on whether you are
using one transducer or two. When using one transducer, Plane 1 and Plane 2
runs are performed sequentially, allowing you to move the transducer between
the planes. When using two transducers, it is assumed the transducer positions
remain fixed and all Plane 1 and 2 runs occur simultaneously.

When you use two transducers with two plane balancing, the measurements are
displayed simultaneously.

IMPORTANT

Anytime during the initial run, you can do the following:

•Press F4 (Esc) to exit the balancing program.

•Press 0 (Shift) and F1 (Go to) simultaneously to go back to a

previous step in the balance run.

See

Move Around in a Balance Run on page 60

for details.

•Press F2 (Summary) to display the results in the Vibration

Summary Table.