2 pin assignment of the "remote" socket – Metrohm 794 Basic Titrino User Manual

Page 165

Advertising
background image

6.2 Pin assignment of the "Remote" socket

794 Basic Titrino

161

6.2 Pin assignment of the "Remote" socket

external

Function

Inputs

pin 21 (Input 0)

pin 9 (Input 1)

pin 22 (Input 2)

pin 10 (Input 3)

pin 23 (Input 4)

pin 11 (Input 5)

pin 24 (Input 6)

pin 12 (Input 7)

Start

Stop

Enter

Clear

Sample

ready

Outputs

pin 5 (Output 0)

pin 18 (Output 1)

pin 4 (Output 2)

Ready

inactive

Conditioning ok,

active if Cond.ok

Titration,

active during titration

Functions see page 163

t

p

> 100 ms

t

p

+5 V

+5 V

+5 V

+5 V

+5 V

+5 V

+5 V

+5 V

+5 V

+5 V

+5 V

not used

Are not used in titration