4 "chloride – Metrohm 794 Basic Titrino User Manual

Page 172

Advertising
background image

6.3. User methods

794 Basic Titrino

168

6.3.4 "Chloride"

Electrode:

6.0430.100 Ag-Titrode at measuring input 1.

Titrant:

c(AgNO

3

) = 0.01 mol/L for low Cl

-

contents,

e.g. in tap water

c(AgNO

3

) = 0.1 mol/L for higher Cl

-

contents,

e.g.

in

food

Sample:

Dissolve sample and add HNO

3

.

Remarks:

Select the appropriate formula. The other may

be deleted.

Reference:

METROHM Application Bulletin Nr. 130

Content of Cl

-

in ppm

Content of table salt in %

Sample size in g

Concentration of titrant

Molar mass of Cl

Factor for ppm

Molar mass of NaCl

Concentration of titrant

Factor for %

'pa

794 Titrino 01102 794.0010

date 2002-01-03 time 09:40 0

DET U Chloride

parameters

>titration parameters

meas.pt.density 4

min.incr. 10.0 µl

titr.rate max. ml/min

signal drift 50 mV/min

equilibr.time 26 s

start V: OFF

pause 0 s

meas.input: 1

temperature 25.0 °C

>stop conditions

stop V: abs.

stop V 99.99 ml

stop U OFF mV

stop EP 1

filling rate max. ml/min

>statistics

status: OFF

>evaluation

EPC 5

EP recognition: all

fix EP1 at U OFF mV

pK/HNP: OFF

>preselections

req.ident: OFF

req.smpl size: all

activate pulse: OFF

------------

'fm

794 Titrino 01102 794.0010

date 2002-01-03 time 09:40 0

DET U Chloride

>calculations

Chloride=EP1*C01*C02*C03/C00;2;ppm

NaCl=EP1*C04*C05*C06/C00;2;%

C00= 1.0

C01= 0.01

C02= 35.45

C03= 1000

C04= 58.44

C05= 0.1

C06= 0.1

------------

Advertising