3 parameters for meas – Metrohm 794 Basic Titrino User Manual

Page 49

Advertising
background image

2.7 Parameters, key <PARAM>

794 Basic Titrino

45

2.7.3 Parameters for MEAS






measuring parameters

determine the measurement.

statistics

:

Calculation of mean values and standard deviation,

see also page 55.

preselections

:

ON/OFF of various auxiliary functions such as auto-

matic requests after the start and activate pulse.


>measuring parameters

signal drift OFF mV/min








equilibr.time OFF s












meas.input: 1






I(pol) 1 µA

U(pol) 400 mV



electrode test: OFF


Measuring parameters

Drift criterion for measured value acquisition (input

range depends on the measured quantity:

pH, U, Ipol:

0.5...999 mV/min, OFF

Upol:

0.05...99.9 µA/min, OFF

T:

0.5...999 °C/min, OFF)

<CLEAR> sets "OFF".

"OFF" means that the measured value is acquired after

a fixed equilibration time.

Waiting time for measured value acquisition (0...9999 s,

OFF)

<CLEAR> sets "OFF".

If no new equilibration time has been entered, the

Titrino calculates an equilibration time appropriate to

the drift with the formula

equilibr.time (in s)

5

0.01

Drift

150

+

+

=

The measured value is acquired when the first criterion

(drift or time) is met. With drift and time "OFF", you will

have an "infinite" measurement.

Measuring input for pH and U. (1, 2, diff.)

Inquiry only with measured quantities pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 155.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 µA)

or the

polarization potential (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive standby mode to a

measurement. "OFF" means that the test is not

parameters

>measuring parameters