2.4. Tutorial

794 Basic Titrino

18

2.4.5 Statistics, acid capacity of drinking water

Let us now determine the acid capacity of drinking water. For this, the SET (set endpoint

titration) mode is used to titrate to pH = 4.3.

First select the mode SET pH (keys <MODE> and <SELECT>).

Now set the end point.

Entry of the end point and the control parameters

<PARAM>

<ENTER>

<ENTER>

<ENTER>

<ENTER>

2 × <QUIT>

• Press <PARAM>. The display shows:

Control parameters for EP1


• Set the end point to pH = 4.3.

• Dynamics means the control range in which the

adding of the titrant is controlled in dependence on

the measured value (OFF: largest control range, i.e.

slow titration).

With buffered systems such as drinking water, this

value can be set lower, e.g. to 3. I.e. 3 pH units be-

fore the set EP controlled dosing starts.

Maximum possible titration rate inside the control

range.

Minimum titration rate inside the control range.


• Quit the inquiry with <QUIT>.

Mount an Exchange Unit with c(HCl) = 0.1 mol/L as titrant. Add 25 mL drinking water as

sample and start the titration with <START>. During the titration the first line of the dis-

play shows the measured value and dispensed volume. The "c" in pHc shows that the

electrode assembly has been calibrated. The second line shows a "control bar", which

indicates the control deviation of the current measured value from the set end point.

parameters

>SET1

>SET1

EP at pH

OFF

>SET1

EP at pH

4.3

>SET1

dynamics OFF

>SET1

dynamics 3

>SET1

max.rate 10.0

ml/min

>SET1

min.rate 25.0µl/min

pHc 6.34

0.426 ml

#======