11 data output – Metrohm 794 Basic Titrino User Manual

Page 62

Advertising
background image

2.11. Data output

794 Basic Titrino

58

2.11 Data output

Reports for the output at the end of a determination







With <DEF>, the report sequence at the end of the

determination is defined.

The entries are specific to the method and are stored in

the method memory.

report:
Definition of report blocks to be printed automatically at
the end of the determination.


The display texts of the Titrino are shown to the left.
The values are the default values.

>report

report:






report:full;curve

Report sequence (input range depends on the mode:

DET: full, short, mplist, curve, derive, comb, scalc full,

scalc srt, param, calc, calib, ff

MET: full, short, mplist, curve, scalc full, scalc srt,

param, calc, calib, ff

SET, MEAS, CAL:

full, short, scalc full, scalc srt, param, calc, calib,

ff

TIP: full, short, scalc full, scalc srt, param, calc, ff

Select a block with <SELECT>. If you require more

than one report block, set a ";" as a delimiter between

the blocks.

Meaning of the report blocks:

full

Full result report with raw results, calculations and statistics.

short

Short result report with calculations and statistics.

mplist

Measuring point list.

curve

Titration curve (with DET and MET) or volume vs. time (with SET) or

measured value vs. time (with MEAS)

derive

1st derivative of the titration curve (with DET)

comb

Combined titration curve and 1st derivative (with DET)

scalc full Full report of silo calculations.

scalc srt

Short report of silo calculations.

param Parameter

report.

calc

Report with formulas and operands.

calib Calibration

data.

ff

Form feed on printer.


DEF

(

2

def

>report

Advertising