6.3 User methods

794 Basic Titrino

167

6.3.3 "Blank"

Treat the "blank sample" in the same way as

you treat your real samples. Can be used e.g.

for the methods Sapon.No. and Perox.No,

see pages 172 and 176.

The blank value is stored as common variable

C30 and can therefore be used directly by

subsequent methods.

Reference:

METROHM Application Bulletin Nr. 210:



















– Result as reagent consumption in mL

– Division factor if multiple quantities of solvent are

used for the blank determination.

'pa

794 Titrino 01102 794.0010

date 2002-01-03 time 09:38 0

DET U Blank

parameters

>titration parameters

meas.pt.density 4

min.incr. 10.0 µl

titr.rate max. ml/min

signal drift 50 mV/min

equilibr.time 26 s

start V: OFF

pause 0 s

meas.input: 1

temperature 25.0 °C

>stop conditions

stop V: abs.

stop V 30 ml

stop U OFF mV

stop EP 9

filling rate max. ml/min

>statistics

status: ON

mean n= 3

res.tab: original

>evaluation

EPC 5

EP recognition: all

fix EP1 at U OFF mV

pK/HNP: OFF

>preselections

req.ident: OFF

req.smpl size: OFF

activate pulse: OFF

------------

'fm

794 Titrino 01102 794.0010

date 2002-01-03 time 09:38 0

DET U Blank

>calculations

Blank=EP1/C01;3;ml

C01= 1

------------