6.3. User methods

794 Basic Titrino

166

6.3.2 "Titer_pH"

The titer is calculated as a factor without di-

mension out of 5 determinations and stored

as common variable C31. It can therefore be

used directly by subsequent methods.

Electrode:

6.0232.100 combined glass pH electrode, at

measuring input 1.

Titrant:

c(NaOH) = 0.1 mol/L (free of carbonate)

Sample:

Weigh in app. 300 mg potassium hydrogen

phtalate (PHP). Dried 2 h at 105 °C, diluted in

40 mL dist. Water (free of carbonate).

Reference:

METROHM Application Bulletin Nr. 206:


– Result (Titer) as a digit without dimension

Sample size PHP in g

– Theoretical consumption for 1 mol PHP

– Molar mass PHP

'pa

794 Titrino 01102 794.0010

date 2002-01-03 time 09:34 0

DET pH Titer_pH

parameters

>titration parameters

meas.pt.density 4

min.incr. 10.0 µl

titr.rate max. ml/min

signal drift 50 mV/min

equilibr.time 26 s

start V: OFF

pause 0 s

meas.input: 1

temperature 25.0 °C

>stop conditions

stop V: abs.

stop V 20 ml

stop pH OFF

stop EP 9

filling rate max. ml/min

>statistics

status: ON

mean n= 5

res.tab: original

>evaluation

EPC 5

EP recognition: all

fix EP1 at pH OFF

pK/HNP: OFF

>preselections

req.ident: OFF

req.smpl size: value

activate pulse: OFF

------------

'fm

794 Titrino 01102 794.0010

date 2002-01-03 time 09:34 0

DET pH Titer_pH

>calculations

Titer=C00*C01/C02/EP1;4;

C00= 1.0

C01= 10000

C02= 204.23

------------