2.14. Calibration data, key <CAL.DATA>

794 Basic Titrino

68

2.14 Calibration data, key <CAL.DATA>





With <CAL.DATA>, the current pH calibration data of

all measuring inputs can be seen. Calibration data are

entered here automatically on completion of a

calibration.

input 1:
Calibration data for measuring input 1.
Identical for input 2 and diff.

The display texts of the Titrino are shown to the left.
The values are the default values.

>input 1

pH(as) 7.00



slope 1.000



temp 25.0 °C



cal.date



electr.id

pH calibration data for measuring input 1

Asymmetry pH (0...±20.00)

Entered automatically after a calibration with measuring

input 1.

Slope (0...±9.999)

Entered automatically after a calibration with measuring

input 1.

Calibration temperature (-20.0...120.0 °C)

Will be printed automatically after calibration with

measuring input 1.

Date of last calibration (no entry possible)

If the calibration data "pH(as)" and/or "slope" are

changed by a manual entry, the date entry is deleted.

Electrode identification of calibrated electrodes (no

entry possible)

If an electrode identification has been entered in the

CAL mode, it is automatically entered after the

calibration.

The calibration report with the current measuring input data can be printed at any time

with the key sequence

<PRINT><CAL.DATA><ENTER>.