6 selection of the mode, key <mode – Metrohm 794 Basic Titrino User Manual
Page 28
2.6. Selection of the mode, key <MODE>
794 Basic Titrino
24
>common variables
C30 0.0
etc.
Values of the common variables
Common variables C30...C39 (0..
±
999 999)
The values of all common variables are displayed. For
creating of common variables see page 57.
2.6 Selection of the mode, key <MODE>
Press key <MODE> until the desired mode is
displayed and confirm with <ENTER>.
Select the measured quantity pH, U, Ipol, Upol, (T) with
<SELECT> and confirm it also with <ENTER>.
The following modes can be selected:
• DET: Dynamic Equivalence-point Titration
• MET: Monotonic Equivalence-point Titration
• SET: Set Endpoint Titration.
• CAL: pH Calibration.
• MEAS: Measuring.
• TIP: Titration Procedure. Linking of various
commands and methods to a titration procedure.
These standard modes are equipped with a set of
standard parameters. They only need few settings in
order to be ready to work.
TIP is an empty "shell". The TIP sequence has to be
defined with <DEF>, see page 62.