5 parameters for tip – Metrohm 794 Basic Titrino User Manual
Page 54
2.7. Parameters, key <PARAM>
794 Basic Titrino
50
2.7.5 Parameters for TIP
In TIP, several commands and methods can be linked to make a titration procedure. The
TIP sequence can be defined with <DEF>, see page 62.
sequence
Parameters for the TIP sequence see page 63.
statistics
:
Calculation of mean values and standard deviation,
see also page 55.
preselections
:
ON/OFF of various auxiliary functions such as auto-
matic requests after the start and activate pulse.
>statistics
status: OFF
mean n= 2
res.tab: original
delete n= 1
Statistics calculation
Status of statistics calculation (OFF, ON)
If the statistics calculation is switched off, the following
inquiries regarding the statistics do not appear.
Mean value calculation from n single results (2...20)
Result table (original, delete n, delete all)
"original":
The original table is used. Deleted
individual results are again incorporated in
the evaluation.
"delete n": Deletion of single results with the index n.
"delete all": The entire table is deleted.
Delete data from sample number n (1...20)
The deleted result is removed from the statistics
calculation.
>preselections
req.ident.: OFF
req.smpl size: OFF
meas.mode: OFF
meas.input: 1
Preselections for the sequence
Request of sample identifications after start (id1, id1&2,
all, OFF)
After start, sample identifications can be inquired
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after the start (value, unit, all,
OFF)
Measured quantity (pH, U, Ipol, Upol, T, OFF)
Quantity for measurements with key <MEAS/HOLD>.
Measuring input (1, 2, diff.)
Inquiry only with measured quantities pH and U.
Measuring input 1 or 2 or differential amplifier;
connection of electrodes, see page 155.
parameters
>sequence