3 assembling the reaction and measuring vessels, Assembling the reaction and measuring vessels – Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 22

Advertising
background image

3.2 Mounting accessories

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

893 Professional Biodiesel Rancimat

Mount the accessories for the external air supply as follows:

1 Mounting the FEP tubing

Screw one end of the FEP tubing 130 mm to the Air/N

2

in con-

nector (2-12) on the rear of the 893 Professional Biodiesel Ranci-
mat.

Screw the tubing adapter M6/olive (6.1808.020) onto the other
end of the FEP tubing.

2 Connecting the gas supply

Mount the compressed air bottle on the M6/olive tubing adapter
(6.1808.020).

Adjust the air flow by means of the reducing valve at the com-
pressed air bottle.

3.2.3

Assembling the reaction and measuring vessels

The following figure shows in detail how the accessory parts for measur-
ing the oxidation stability have to be mounted and connected to one
another.