Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 35

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

893 Professional Biodiesel Rancimat

■■■■■■■■

27

Preparing the determination of the temperature correction

The figure shows in detail how the accessory parts are to be assembled
for the determination of the temperature correction. Proceed as follows:

1 Preparing the reaction vessel cover

Mount the air tube on the reaction vessel cover.

Clamp the first spacer at a distance of approx. 12 cm from the
lower end onto the air tube.

Clamp the second spacer at a distance of approx. 5 cm from the
lower end onto the air tube.

Insert the temperature sensor from above into the temperature
sensor opening and fasten it in the corresponding openings of the
spacers.

2 Preparing the reaction vessel

Fill the reaction vessel with 6 g silicone oil (6.2326.000).

Place the reaction vessel cover with the temperature sensor on
the reaction vessel.

Push the temperature sensor all the way down (the sensor must
touch the vessel base).

3 Inserting and connecting the reaction vessel

Insert the reaction vessel with the mounted reaction vessel cover
in the measuring position 2 or 3 of the required heating block.

Screw one end of the FEP tubing 250 mm to the M8/M6 thread
adapter of the reaction vessel cover.

Screw the other end of the FEP tubing to the corresponding con-
nector of the 893 Professional Biodiesel Rancimat.

Connect the temperature sensor to the Pt100 connector (2-14)
on the rear of the 893 Professional Biodiesel Rancimat.

NOTE

You will find detailed information regarding the StabNet computer
program in the Tutorial.

Advertising