Figure 8, Mounting the air tube: correct - incorrect, Figure 9 – Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 25: Mounting the foam barrier

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

893 Professional Biodiesel Rancimat

■■■■■■■■

17

Figure 8

Mounting the air tube: correct - incorrect

WARNING

The foam barrier can melt if it projects too deeply into the heating
block.

Ensure that the foam barrier is at least 7 cm above the base of the
reaction vessel.

Figure 9

Mounting the foam barrier

3.2.4

Inserting vessels / Establishing tubing connections

After you have assembled the reaction and measuring vessels, insert them
in the 893 Professional Biodiesel Rancimat and establish the tubing con-
nections (see Chapter 2.1, page 6) as follows:

1 Inserting the measuring vessel

Fill distilled water into the measuring vessel.

Place the measuring vessel cover onto the measuring vessel.

Insert the measuring vessel into the openings provided on the 893
Professional Biodiesel Rancimat. While doing so, carefully guide
the connector plug into the electrode connector.

Connect the black Iso-Versinic

®

tubing to the M8/olive tubing

adapter of the measuring vessel cover.