3.2 Mounting accessories

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

893 Professional Biodiesel Rancimat

2 Mounting the tubing for the air supply

Screw the FEP tubing 250 mm to the air supply connections of
the 893 Professional Biodiesel Rancimat.

3 Inserting the reaction vessel

Fill the reaction vessel with the sample.

After the required reaction temperature has been reached, insert
the reaction vessel with the mounted reaction vessel cover in the
openings provided on the 893 Professional Biodiesel Rancimat.

4 Establishing the tubing connections

Connect the Iso-Versinic

®

tubing that is connected to the measur-

ing vessel cover to the tubing connector of the reaction vessel
cover.

Screw the FEP tubing 250 mm which is connected to the M8/olive
tubing adapter of the 893 Professional Biodiesel Rancimat to the
M8/M6 thread adapter of the reaction vessel cover.

3.2.5

Mounting the exhaust air collection tube

The optional exhaust air collection tube (6.2757.000)

[Link target not

found in publication context!]

can be mounted on the 893 Professional

Biodiesel Rancimat for targeted removal of the exhaust air.

NOTE

A total of 8 pieces of Iso-Versinic

®

tubing (6.1839.000) are required in

addition to the exhaust air collection tube.

Proceed as follows to mount the collection tube:

1 Mounting the exhaust air collection tube

Insert the exhaust air collection tube with both support pins into
the collection tube holders on the 893 Professional Biodiesel Ran-
cimat in such a way that the connector to the exhaust air removal
is located to the rear.

2 Connecting the measuring vessels

Screw the M8/olive tubing adapter into the Out opening of the
measuring vessel cover.

Connect one end of the Iso-Versinic

®

tubing to the M8/olive tub-

ing adapter.