7 technical specifications, 1 general data, 2 temperature control and measurement – Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 51: General data, Temperature control and measurement

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Technical specifications

893 Professional Biodiesel Rancimat

■■■■■■■■

43

7 Technical specifications

7.1

General data

Number of instru-
ments that can be
connected

1 - 4

Number of sam-
ples per instru-
ment

1 - 8 (4 each per heating block)

Sample amount

Several grams / several milliliters

7.2

Temperature control and measurement

Sample tempera-
ture setting range

50 - 150 °C

Temperature cor-
rection

±0 - 9.9 °C
(Can be entered manually or determined automatically using the exter-
nal temperature sensor.)

Temperature mea-
suring range

0 - 300 °C

Resolution

0.1 °C

Maximum devia-
tion of the heating
block temperature
from the set value
(50 - 220 °C)

±0.3 °C

Reproducibility of
the set tempera-
ture

typ. ±0.2 °C

Max. temperature
difference
between different
measuring posi-
tions per block

typ. < 0.3 °C

Temperature fluc-
tuations

typ. < 0.1 °C
(With operating temperature attained, reaction vessels inserted and
identically filled and 20 L/h air throughput.)

Advertising