7 technical specifications, 1 general data, 2 temperature control and measurement – Metrohm 893 Professional Biodiesel Rancimat User Manual
Page 51: General data, Temperature control and measurement
■■■■■■■■■■■■■■■■■■■■■■
7 Technical specifications
893 Professional Biodiesel Rancimat
■■■■■■■■
43
7 Technical specifications
7.1
General data
Number of instru-
ments that can be
connected
1 - 4
Number of sam-
ples per instru-
ment
1 - 8 (4 each per heating block)
Sample amount
Several grams / several milliliters
7.2
Temperature control and measurement
Sample tempera-
ture setting range
50 - 150 °C
Temperature cor-
rection
±0 - 9.9 °C
(Can be entered manually or determined automatically using the exter-
nal temperature sensor.)
Temperature mea-
suring range
0 - 300 °C
Resolution
0.1 °C
Maximum devia-
tion of the heating
block temperature
from the set value
(50 - 220 °C)
±0.3 °C
Reproducibility of
the set tempera-
ture
typ. ±0.2 °C
Max. temperature
difference
between different
measuring posi-
tions per block
typ. < 0.3 °C
Temperature fluc-
tuations
typ. < 0.1 °C
(With operating temperature attained, reaction vessels inserted and
identically filled and 20 L/h air throughput.)