7.3 External temperature sensor

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

893 Professional Biodiesel Rancimat

Shutdown temper-
ature

180 ± 7 °C
(Resetting and troubleshooting carried out by Metrohm Service.)

Heating time for
the instrument

approx. 45 min (from 20 °C to 120 °C)

External tempera-
ture of the instru-
ment

< 50 °C
(at operating temperature 150 °C)

7.3

External temperature sensor

Sensor

4-pin for sensor Pt100 (6.1111.020)

7.4

Conductivity measurement

Sensor

Conductivity measuring cell, integrated in the measuring vessel cover
(6.0913.130)

Electrode

Conductivity measuring cell with 2 stainless steel electrodes

Measuring princi-
ple

AC current measurement with 1 kHz frequency and approx. 1.0 V
amplitude (peak to peak)

Cell constant

1.0 - 1.2
(The exact value can be entered manually or determined automatically.)

Measuring range

0 - 400 µS/cm (at c = 1)

Resolution

0.1 µS/cm

Display range

0 - 999 µS/cm

Maximum devia-
tion from mea-
sured value

±(0.5 µS/cm + 1% of the measured value)