4 cleaning the oil separator, Cleaning the oil separator – Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 40

Advertising
background image

4.3 Determinations

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

893 Professional Biodiesel Rancimat

Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and
thoroughly rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Remove the protective ring to facilitate cleaning of the electrodes.

2 Cleaning the reaction vessels and accessories

Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.

Clean the reaction vessel covers with acetone or 2-propanol or
in the dishwasher and rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

Then heat the reaction vessel covers in the drying oven for 2
hours at 80 °C.

NOTE

Replace the reaction vessel covers if they no longer sit tightly on
the reaction vessel or the material is brittle.

3 Cleaning the tubing

Clean the Iso-Versinic

®

tubing with acetone or 2-propanol or in

the dishwasher and rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

Then heat the Iso-Versinic

®

tubing in the drying oven for 2

hours at 80 °C.

4.3.4

Cleaning the oil separator

The oil separator is available as an optional accessory.