background image

4.2 Calibration functions

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

893 Professional Biodiesel Rancimat

Figure 14

Measuring arrangement (schematic representation)

1

Reaction vessel

2

Sample

3

Heating block

4

Measuring vessel

5

Conductivity measuring cell

6

Measuring solution

4.2

Calibration functions

4.2.1

Determining the cell constant

The cell constant of the conductivity sensor (6.0913.130) is normally at
1.1 ± 0.1 cm-1. This accuracy is sufficient for an induction time determi-
nation since only the shape of the curve is evaluated. For the determina-
tion of the stability time, however, the conductivity change is measured
in absolute terms. As a rule, the time until an increase in conductivity of
50 µS/cm is determined. The cell constant for the conductivity electrode
being used has to be calibrated in order to be able to measure the con-
ductivity correctly.