Metrohm NIRS DS2500 Analyzer User Manual

Page 22

Advertising
background image

3.6 Initial start-up

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

3

In the software, start Reference Standardization. Follow the
instructions in the software.

When the software prompt appears, insert the reflection standard
(R8012007) into the opening of the holder. The label of the
reflection standard must be visible.

Close the lid.

Wait until Reference Standardization is completed.

Remove the reflection standard from the holder.

Regulated industries

If you put the NIRS DS2500 Analyzer into operation in a regulated indus-
try, then additional tests are required after the procedures described in the
chapter "Non-regulated industries" have been performed.

In regulated industries, the Vision Pharma software version and the NIRS
reflection standard, 7 pcs. (6.7450.010) are required for the instrument to
be put into operation.

Information on which additional tests are required for your specific regu-
lated industry can be found in the applicable standards.