3 inserting the sample cup for liquid samples, Inserting the sample cup for liquid samples – Metrohm NIRS DS2500 Analyzer User Manual
Page 25
Advertising
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
■■■■■■■■
17
3.7.3
Inserting the sample cup for liquid samples
Inserting the sample cup for liquid samples
Required accessories
■
NIRS liquid sample kit transflection (6.7400.010)
■
NIRS DS2500 holder for sample cup (6.7430.040)
1 Clean the sample cup and the sampling window with a lens cleaner
cloth.
2
■
Fill liquid sample to a height of approx. 1 cm into the sample cup.
■
Place the gold diffuse reflector into the liquid. Avoid air pockets
while doing so.
3
■
Place the sample cup holder onto the sampling window.
The metal pin on the holder must go into the indentation on the
sampling window.
■
Place the sample cup into the round opening of the holder.