3.7 Setting up accessories

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

3.7.2

Inserting the sample cup for powders

Use the NIRS mini sample cup with disposable back (6.7402.030) for mea-
suring fine solid samples (powder).

Inserting the sample cup for powders

Required accessories

NIRS DS2500 mini sample cup (6.7402.030)

NIRS DS2500 holder for sample cup (6.7430.040)

1 Clean the sample cup and the sampling window with a lens cleaner

cloth.

2

Fill the powder into the sample cup.
Make sure that the glass bottom of the sample cup is completely
covered and that the layer of sample is at least 1 cm thick.

Seal the sample cup with the disposable back.

3

Place the sample cup holder onto the sampling window.
The metal pin on the holder must go into the indentation on the
sampling window.

Place the sample cup into the round opening of the holder.