6 Maintenance

■■■■■■■■

27

4 Inserting the lamp holder

First, push the white lamp holder carefully all the way into the
opening,

then turn it approx. 45° clockwise and

then slowly let it go.

5 Closing the lamp compartment

Place the sealing plate onto the opening. Make sure that no
cables are being pinched.

Insert the four screws and tighten them in crosswise sequence
using the hex key.

Push the hex key into the guide rail on the cover.

Close the lid.

6 Calibrating the instrument

The instrument has to be recalibrated every time a lamp is replaced.

Switch on the instrument.

Recalibrating the instrument (see step 6 in the chapter "Non-reg-
ulated industries", page 12)
.