6.1 Maintenance by Metrohm Service

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

6 Maintenance

6.1

Maintenance by Metrohm Service

Maintenance of the NIRS DS2500 Analyzer is best carried out as part of
annual service, which is performed by specialist personnel from Metrohm.
A shorter maintenance interval may be necessary if you frequently work
with caustic and corrosive chemicals.

Metrohm Service offers every form of technical advice for maintenance
and service of all Metrohm instruments.

6.2

User maintenance

Table 1

Maintenance tasks and intervals

Maintenance task

Maintenance interval

Cleaning the instrument

when necessary

Cleaning the sample compartment

when necessary

Cleaning the sampling window

before every measurement

Replacing the lamp

after 4,000 burning hours

if Wavelength Certification or the Performance
Test
deliver unsatisfactory results.

Check the fan filter and replace it, if necessary

monthly

Replacing the fuse

if melted

6.2.1

Replacing the lamp

The lamp must be replaced if it is burned out or when its performance is
insufficient. Signs of an insufficient lamp performance include:

Measurements are impaired by noise.

The repeatability of the wavelengths is deteriorating.

The performance test is no longer completed successfully.

Removing the lamp

Required accessories

Screwdriver

Clean the instrument before replacing the lamp. This prevents dust caus-
ing damage to the lamp's reflector.