6.2 User maintenance

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

6.2.2

Replacing the fan filter

The fan filter has to be checked at least once a month. If the instrument is
operated in a dusty or otherwise dirty environment, then a check is
required once or even twice a week.

The fan is located on the rear of the instrument. The filter cover is
attached to the fan filter by means of four plastic latches.

Required accessories

Fan filter (6.7430.020)

1 Switching off the instrument

Turn the on/off switch (2-1) to the position O.

2 Removing the filter cover

Grab the filter cover with both hands and gently take it off starting
on top and then proceeding to the sides.

3 Checking the state of the filter

Take out the filter and inspect it carefully.

If the filter shows no cracks, then it can be cleaned and used
again.

If the filter is damaged, then a new filter has to be used.