6.2 User maintenance

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

6.2.3

Replacing the fuses

The fuse is located on the rear of the instrument, directly underneath the
on/off switch.

Required accessories

Spare fuse, type: 250 V, 5 A, slow-acting fuse, 20 mm

1 Switching off the instrument

Turn the on/off switch (2-1) to the position O.

NOTE

We recommend pulling out the power supply cable in addition.
The instrument can thus not be accidentally switched on while you
insert the fuse.

2 Removing the old fuse

Unscrew the fuse holder by hand (if necessary, unscrew it using a
size 5 flat-head screwdriver).

Pull the old fuse out of the holder.