2 mounting base tray and bottle holder (optional), 1 removing/mounting the base tray, Mounting base tray and bottle holder (optional) – Metrohm 941 Eluent Production Module User Manual

Page 19

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

941 Eluent Production Module (2.941.0010)

■■■■■■■■

11

The bottle holder and base tray must be removed or set up every time one
of the following instruments is to be mounted on or under a 940 Profes-
sional IC Vario:

One or more 942 Extension Module Vario.

Or another instrument with the same-sized footprint.

4.2.2

Mounting base tray and bottle holder (optional)

The base tray and bottle holder come fully assembled on a new ion chro-
matograph. If you wished to install an extension module on the ion chro-
matograph, you would have to remove the bottle holder and put it back
on top of the topmost instrument. If you wished to install an extension
module below the ion chromatograph, you would have to remove the
base tray and set it under the lowest instrument.

4.2.2.1

Removing/mounting the base tray

The base tray must be removed if you want to install another instrument
under the IC instrument.

CAUTION

Do not allow capillaries or leak sensor cables to be pinched!

Pinches in the leak sensor cable or the capillaries fed through the guide
ducts between the base tray and the instrument may lead to malfunc-
tions.

Unplug the leak sensor cable. Remove all of the capillaries from the
capillary ducts.

Removing the base tray

Before you can remove the base tray, the following preconditions must be
met:

The instrument is switched off.

The bottle holder is cleared.

All of the cable connections on the rear have been disconnected.

The capillaries are removed from the guide ducts between the instru-
ment and the base tray.

There are no loose parts in the instrument.

To remove the base tray, you need a 3 mm hex key (6.2621.100).