Metrohm 941 Eluent Production Module User Manual

Page 29

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

941 Eluent Production Module (2.941.0010)

■■■■■■■■

21

3 Sealing the M8 opening

Screw the M8 threaded stopper into the M8 opening.

4 Connecting the level sensor cable

Connect the two plugs of the level sensor cable to the two sockets.

5 Mounting the adsorber

Fill the adsorber tube with cotton and insert it into the last remaining
opening.

6 Bundling the level sensor cables and tubings

NOTE

The fine connection cables of the level sensor cable break off easily
if the bottle cap is rotated too much.

The level sensor cable is damaged as a result and needs to be
replaced.

To prevent the connection cables from breaking off, we recom-
mend bundling the level sensor cable together with the tubings
and capillaries that are connected to the bottle cap.

Tie the level sensor cable and the FEP tubing together using a spiral
band. Tie a piece of spiral band with a length of approx. 10 cm
around the cable and the tubings as close as possible to the bottle
cap.

7 Attaching the bottle cap

Screw the bottle cap onto the concentrate bottle.

Advertising