4 connecting the concentrate bottle, Connecting the concentrate bottle, Figure 5 – Metrohm 941 Eluent Production Module User Manual

Page 27

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

941 Eluent Production Module (2.941.0010)

■■■■■■■■

19

4.4

Connecting the concentrate bottle

The bottle with the concentrate can be equipped with a bottle cap with-
out level sensor, e.g. with the bottle cap (6.1602.160).

Accessories

For this step you need the following accessories:

Bottle (6.1608.070) filled with concentrate

Bottle cap (6.1602.160)

FEP aspiration tubing (6.1819.100)

FEP tubing (6.1805.530)

Adsorber tube (6.1609.000)

SGJ clip for SGJ 14/15 (6.2023.020)

Figure 5

Installing the concentrate bottle cap without level sensor

1 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening.

Insert the FEP aspiration tubing into the M6 opening.

Screw the FEP tubing into the M6 opening.

2 Sealing the M8 opening

Screw the M8 threaded stopper into the M8 opening.

3 Mounting the adsorber

Fill the adsorber tube with adsorber material and cotton (for anion
eluents) or only with cotton (for cation eluents) and insert it into the
large opening. Secure in place with an SGJ clip.