4.5 Connecting the ultrapure water container

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

941 Eluent Production Module (2.941.0010)

2 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening.

Insert the FEP aspiration tubing (6.1819.110) into the M6 open-
ing.

Screw the FEP tubing into the M6 opening.

3 Sealing the M8 opening

Seal the M8 opening with a stopper.

4 Connecting the level sensor cable

Connect the two plugs of the level sensor cable to the two sockets.

5 Mounting the adsorber

Fill the adsorber tube with soda lime and cotton and insert it with the
adapter into the last remaining opening.

6 Bundling the level sensor cables and tubings

NOTE

The fine connection cables of the level sensor cable break off easily
if the bottle cap is rotated too much.

The level sensor cable is damaged as a result and needs to be
replaced.

To prevent the connection cables from breaking off, we recom-
mend bundling the level sensor cable together with the tubings
and capillaries that are connected to the bottle cap.

Tie the level sensor cable and the FEP tubing together using a spiral
band. Tie a piece of spiral band with a length of approx. 10 cm
around the cable and the tubings as close as possible to the bottle
cap.

7 Attaching the bottle cap

Screw the bottle cap onto the canister.