Figure 4, Installing the eluent bottle cap with level sensor – Metrohm 941 Eluent Production Module User Manual

Page 25

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

941 Eluent Production Module (2.941.0010)

■■■■■■■■

17

3 Insert the two measuring sensors from the inside into the two open-

ings for measuring sensors.

Assembling the bottle cap

Accessories

For this step you need:

Eluent bottle (6.1608.070) from the accessories set of the ion chroma-
tograph.

Bottle cap for level sensor (6.1626.000) with inserted measuring sen-
sors (see "Inserting the measuring sensor", page 16).

FEP aspiration tubing (6.1819.100)

FEP tubing (6.1805.530)

Level sensor cable (6.2151.060)

Adsorber tube (6.1619.000)

Spiral band (6.1815.010)

Figure 4

Installing the eluent bottle cap with level sensor

1 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening.

Insert the FEP aspiration tubing into the M6 opening.

Screw the FEP tubing into the M6 opening.