4.5 Connecting the ultrapure water container

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

941 Eluent Production Module (2.941.0010)

4.5

Connecting the ultrapure water container

The ultrapure water can be aspirated either from a 10 L canister or an
ELGA PURELAB water purification system.

The 10 L canister is included in the scope of delivery and is used for the
standard installation.

4.5.1

Connecting the ultrapure water canister

Assembling the bottle cap without level sensor
The canister with the ultrapure water can be equipped with a bottle cap
without level sensor, e.g. with the bottle cap (6.1602.160).

Accessories

For this step you need the following accessories:

Canister, 10 L (6.1621.000)

Bottle cap (6.1602.160)

FEP aspiration tubing (6.1819.100)

FEP tubing (6.1805.530)

Adsorber tube (6.1609.000 + 6.1624.000)

Figure 7

Installing the ultrapure water canister cap without level sen-
sor

1 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening.

Insert the FEP aspiration tubing into the M6 opening.

Screw the FEP tubing into the M6 opening.