Troubleshooting

7-9

Problem

The velocity data readings are abnormally high.

Possible Cause

Wires on the P2 connector may be loose. The velocity sensor may be covered with silt.
The velocity sensor may be out of the flow. The flow may be reversed or slower than 0.15
m/s (0.5 ft/s).

Solution

Check the site conditions and relocate the velocity sensor if necessary.

If the problem is still unresolved, replace the monitor and send the old one to ADS for
service.

Pressure Depth Subsystem

Use the following tips to help resolve problems with the pressure depth subsystem.

Warning:

Prior to replacing a sensor, contact an ADS trained technician for

further diagnosis.

Problem

The temperature reading from the pressure depth sensor shows a value of -273

°

C

(-459

°

F).

Possible Cause

The cable connecting the pressure mouse to the monitor may be broken. The pressure
sensor may be faulty. The pressure depth board is not responding or may be faulty.

Solution

Replace the pressure sensor with a known good sensor and send the faulty one to ADS for
repair.

If the problem is still unresolved, replace the monitor and send the old one to ADS for
service.

Problem

The pressure depth subsystem data readings are consistently incorrect.

Possible Cause

The coefficients are incorrect.

Solution

Enter the correct coefficients and reactivate the monitor.

Problem

The pressure depth subsystem data readings are consistently off by up to 7.62 cm (3 in.),
but temperature readings are accurate.

Possible Cause

The pressure sensor has an offset that drifts over time.

Solution

Check the condition of the pressure dryer tube.

Adjust the depth data by applying a delta offset to the readings until they match the values
obtained during confirmation.

Problem

The pressure sensor temperature readings are incorrect, but depth data is accurate.

Possible Cause

The temperature sensor may be faulty.

Solution

Contact an ADS trained technician for further information about replacing the pressure
sensor.

If the problem is still unresolved, replace the monitor and send the old one to ADS for
service.