ADS 3600 Flow Monitor O&M Manual

l

Mount the anchor bolt.

l

Place a 7/16 in. (11 mm) nut on the bolt and twist until it stops (hand tight
only). This nut is referred to as the adjustment nut.

l

Replace the plate over the anchor bolt and install a washer and a second 7/16
in. (11 mm) securing nut just enough to hold the plate on the stud.

l

Orient the attached ultrasonic plate in the desired position and press against the
crown (soffit) of pipe.

l

Spot drill (mark) the remaining three holes for the ultrasonic sensor mounting
plate.

l

Swivel the plate 180

°

.

01-05.pcx

Plate Rotation

l

Mount the three anchor bolts as described earlier.

l

Swivel the plate back to correct the position and align the holes.

l

Firmly push the plate up to the adjustment nuts and secure the plate with 7/16
in. (11 mm) securing nuts until they’re tight.

Note:

In some instances the anchor bolt may be slightly offset from the plate

holes. Re-alignment can be done by simply loosening the 7/16 in. (11 mm)
adjustment nut at the end of the anchor bolt and gently tapping with a hammer
until the correct alignment is attained. Tapping the nut eliminates any thread
damage to the anchor bolt.

l

Level the plate with a carpenter’s (torpedo or line) level. A fine-tune
adjustment is done using the adjustment nuts on the back side or upper surface
of the ultrasonic plate.

Note:

In some instances, the plate may bind and can be lowered with the aid

of a flat head screw driver.