Jumper settings – ADS Environmental Services Model 3600 530002 A2 User Manual

Page 92

Advertising
background image

5-18

ADS 3600 Flow Monitor O&M Manual

1

2

3

4

5

6

7

8

Power

Time-out

Spare

106/103

Connection Type

Baud Rate

Open

Monitor will go to sleep.

Closed

Continuous power operation

Open

Time-out enabled (10 min)

Closed

Time-out disabled

Open

N/A

Closed

N/A

Open

103100 CPU

Closed

106100 CPU

Internal

Auto

Answer

Modem

Open

Open

External

Non-auto

Answer

Modem

Closed

Open

Direct

Closed

Closed

RS-232

1200

Open

Open

2400

Closed

Open

4800

Open

Closed

9600

Closed

Closed

SW3

Options Switch (Standard switch settings are shown in shaded rows.)

Jumper Settings

The following table shows the correct jumper settings for the 3600 flow monitor.
However, the only configuration that you may need to change is the battery backup
configuration. If you disable a board and place it in storage, the backup battery will
need to be re-configured prior to use. Only an ADS certified technician can disable
the board.