Series 7 Digital Force Gauges

User’s Guide


15

9.5.1 External Trigger Schematic Diagram

14

GND

1

IN

22k

SWITCH UNDER TEST

GAUGE

+3.3V

TO MICROPROCESSOR

0.1 uF

GND

BAT54SLT1G

330 Ohm

(NO OR NC)

USER - SUPPLIED


Note: Custom cabling is required to connect to a switch, or to connect a switch and a Mark-10 test stand
simultaneously.

9.6 Data Capture (CAPT)
This mode of operation is used to capture and store continuous data in the gauge’s memory. The capture
frequency can be adjusted to accommodate quick-action as well as longer duration tests. Saved data can
be downloaded in bulk via USB or RS-232.

9.6.1 Configuration
After Data Capture has been enabled, it may be selected by pressing the MODE key until CAPT is
displayed. The display appears as follows:


Note: For best performance, it is recommended to keep the current reading filter at its lowest value. See
Digital Filters section for details.

9.6.2 Start Condition
Data capture is initiated when the Start Condition has been triggered. Several triggers are available, as
shown below:

Function Description

Enabled

If enabled, CAPT appears as one of the operating modes.

Period

The capture period may be adjusted by pressing the

and

keys to change

the value of the hours (H), minutes (M), seconds (S), and fractions of seconds (x)
fields. Press the key to advance to the next field.
Available settings: Hours: 0-24, Minutes: 0-59, Seconds: 0-59,
Fraction of Seconds: 0.00007-0.99995, in 0.00007 (70 μS) increments.

Start Condition

See following sub-sections for details.

Auto Settings

DATA CAPTURE

*

Enabled

Period (H:M:S.x)
00 :00 :00 .00007
+ Start Condition
+ Stop Condition
+ Auto Settings