Series 7 Digital Force Gauges

User’s Guide


26

** The output assignments depend on several factors described in the table below. Output functions
always reference the primary reading on the display, regardless of the current mode.

Force

Pin 11

(SP1)

Pin 12

(SP2)

Pin 13

(SP3)

Upper and Lower Set Points are Compression

Greater than or equal to upper set point

On

Off

Off

Between upper and lower set points

Off

Off

On

Less than or equal to lower set point

Off

On

Off

Upper and Lower Set Points are Tension

Greater than or equal to upper set point

Off

On

Off

Between upper and lower set points

Off

Off

On

Less than or equal to lower set point

On

Off

Off

Upper Set Point is Compression, Lower Set Point is Tension

Greater than or equal to upper set point, in compression

Off

On

Off

Between upper and lower set points

Off

Off

On

Greater than or equal to lower set point, in tension

On

Off

Off

Upper Set Point is Tension, Lower Set Point is Compression

Greater than or equal to upper set point, in tension

Off

On

Off

Between upper and lower set points

Off

Off

On

Greater than or equal to lower set point, in compression

On

Off

Off


14.6 Command Set / Gauge Control Language 2 (GCL2)
Series 7 force gauges may be controlled by an external device through the RS-232 or USB channels. The
following is a list of supported commands and their explanations. All commands must be terminated with
a Carriage Return character or with a Carriage Return/Line Feed combination. The gauge responses are
always terminated with a Carriage Return/Line Feed.

Request

Readings

?

Request the displayed reading (dependant on operating mode)

?C

Request the current (real time) reading

?PT

Request the peak tension reading (corresponds to static COF)

?PC

Request the peak compression reading

?ET

Request the reading obtained during the External trigger mode

?A

Request the average reading obtained during the Average mode

(corresponds to kinetic COF)

?P1 Request

1

st

peak reading


Units

LB

Switch units to pound-force

OZ

Switch units to ounce-force

KG

Switch units to kilogram-force

G

Switch units to gram-force

N

Switch units to Newtons

MN

Switch units to Milli-Newtons

KN

Switch units to Kilo-Newtons

COF

Switch units to Coefficient of Friction

CU

Switch to user-defined custom unit

COFE

Enable Coefficient of Friction unit

COFD

Disable Coefficient of Friction unit

CUE

Enable user-defined custom unit

CUD

Disable user-defined custom unit