Cooper Instruments & Systems Series M7 Professional Digital Force Gauge User Manual

Page 17

Advertising
background image

Series 7 Digital Force Gauges

User’s Guide


16


Select “+ More” for additional options:


9.6.3 Stop Condition
Data capture is automatically terminated when the Stop Condition has been reached. Several conditions
are available, as shown below:


When data capture has stopped, the data may be automatically transmitted in bulk to a PC program such
as MESUR

TM

gauge (see the following sub-section for details). Data may also be transmitted manually

through the Memory section.

Note 1: If the Stop Condition does not occur, data capture stops automatically when the memory
is full.

Function

Data capture is initiated when:

Momentary HiLo

An external trigger occurs.

See External Trigger sub-section for details for each of these functions.

Momentary LoHi
Maintained High
Maintained Low
Start Force

The desired trigger force is reached. Toggle between compression and tension
directions by pressing the DIRECTION key.

DATA Key

The DATA key is manually pressed.

Function

Data capture terminates when:

Samples

The desired number of samples (data points) has been captured.

Stop Force

The desired force has been reached.

Memory Full

5,000 data points have been captured.

DATA Key

The DATA key is manually pressed.

START CONDITION 2

Start Force

3.500 lbF
* DATA Key

STOP CONDITION


*

Samples:

10

Stop Force

2.500 lbF

Memory Full
DATA Key

START CONDITION


Momentary HiLo
Momentary LoHi
* Maintained High
Maintained Low
+ More