9.7.2 Peak Settings
The display appears as follows:


Thresholds and percentage drops are illustrated below:

Enabled

If enabled, 2PK appears as one of the operating modes. In the main display, the
Peak readings will reference the first and second peaks – first peak on top,
second peak below. Refer to the Home Screen & Controls section for details.

Peak Settings

See following sub-sections for details.

Auto Output
Auto Store PK1

Automatically stores the first peak reading to memory.

Auto Store PK2

Automatically stores the second peak reading to memory.

Auto Zero

Automatically zeroes the display following data transmission and/or storage.

Threshold 1

Sets the percentage of full scale at which the first/second peak detection
feature becomes active. This threshold is provided to ignore peaks that can
occur during sample loading and unloading. Available settings: 1–90%, in
1% increments between 1–5%, and in 5% increments between 5–90%.

% (Percentage) Drop 1 Identifies the first peak via detection of a specified percentage drop from

peak. Available settings: 5–95%, in 5% increments.

Threshold 2

Same as with Threshold 1, but refers to a percentage increase above the
first peak.

% (Percentage) Drop 2 Same as Percentage Drop 1 but for the second peak.
Auto Zero Delay

Sets the time delay before the peak readings are zeroed. Available settings:
1–10 sec. in 1 sec. increments, and 10–60 sec. in 5 sec. increments.

PEAK SETTINGS

Threshold 1:

5 %

% Drop 1: 10 %
Threshold 2: 5 %
% Drop 2: 10 %
Auto Zero Delay

3 sec.