Series 7 Digital Force Gauges

User’s Guide


19

9.7.3 Auto Output Settings
Scroll to Auto Output Settings and press ENTER to select the desired automatic outputs. Select RS-
232/USB and/or Mitutoyo outputs, and select First and/or Second peaks, and an Output Pin, if desired.
The display appears as follows:

Parameter Description

RS232/USB Output

Automatically output the selected peak(s) after the 2

nd

peak is detected.

Mitutoyo Output

Automatically output the selected peak(s) after the 2

nd

peak is detected.

First Peak

Automatically output the first peak.

Second Peak

Automatically output the second peak.

Output Pin

Automatically toggle the SP1, SP2, or SP3 pins (active low). If not required, select
“NONE”. To stop a Mark-10 ESM301 test stand when a 1

st

/2

nd

peak capture

sequence is completed, specify SP2 if moving in the tension (upwards) direction,
or SP1 for the compression (downwards) direction.
Note: Because the same pins are also used for set points, the stand will stop at
whichever condition occurs first.



10 DATA MEMORY AND STATISTICS

Series 7 gauges have storage capacity of 5,000 data points. Readings may be stored, viewed, and output
to an external device. Individual, or all data points may be deleted. Statistics are calculated for the data in
memory.

To enable memory storage, select DATA Key from the menu, then scroll to Memory Storage and press
ENTER. Then exit the menu. In the home screen, the data record number 0000 appears below the
primary reading. Press DATA at any time to save the displayed reading. The record number will
increment each time DATA is pressed. If DATA is pressed when memory is full the message “MEMORY
FULL” will be flashed at the bottom of the display and a double audio tone will be sounded.

To view, edit, and output stored readings and statistics, select Memory from the menu. The screen
appears as follows:


10.1 View Data
All the saved data points may be viewed. The record number is displayed, along with the corresponding
value and currently set unit of measurement. Any readings may be deleted individually. To do so, scroll to

AUTO OUTPUT

SETTINGS

RS232/USB Output

Mitutoyo Output
First Peak
Second Peak
Output Pin: NONE

MEMORY

View Data

View Statistics
Output Data
Output Statistics
Output Data & Stats
Clear All Data