Series 7 Digital Force Gauges

User’s Guide


27



Basic Functions

CUR

Current mode (real time mode) for primary reading

PT

Peak Tension mode for primary reading

PC

Peak Compression mode for primary reading

CLR

Clear peaks

Z

Zero display and perform the CLR function

Filters

FLTCn

Digital filter for displayed readings

FLTPn

Digital filter for current readings

n= 0-10, filter = 2

n

, ex: n=0= no filter, n=10=1024 samples averaged


Memory & Statistics

MEM

Transmit all stored readings

STA

Transmit

statistics


Set

Points

SPHD

Disable high set point

SPLD

Disable low set point

SPHn

High set point. n=value (+ for compression, - for tension)

SPLn

Low set point. n=value (+ for compression, - for tension)
Note: High set point value must be greater than low set point value if both values
are set to the same polarity.

USB/RS-232

Communication

FULL

USB/RS-232 transmission with units

NUM

USB/RS-232 transmission without units (only numeric values)

AOUTn

Auto-transmit n times per second n=1, 2, 5, 10, 25, 50, 125, 250. 0=disabled
Note: n = 1 = yields 50 times per second. This is provided for backward
compatibility with legacy gauges.

IPOLn

Invert polarity of output. n=1=invert polarity. n=0=normal (default)

Note: Normal polarity is positive for compression and negative for tension.

OPOLn

Omit polarity of output. n=1=omit polarity. n=0=include polarity (default)

Note: The “+” sign is always omitted. A “–“ sign is sent when polarity is enabled.

Mitutoyo Communication

MIT

Enable Mitutoyo output

MITD

Disable Mitutoyo output

POL

Mitutoyo output with polarity (+ for compression, - for tension)

NPOL

Mitutoyo output without polarity (absolute value)

PM

Print/send data to a Mitutoyo-compatible device

Averaging

A

Enable Average mode

AD

Disable Average mode

AM

Select Average mode (if enabled) for primary reading

TRFn

Trigger force. n=value (+ for compression, - for tension)

DELn

Initial delay. n=0.1-300.0 seconds

ATn

Average time. n=0.1-300.0 seconds


External Trigger

ETH

Enable high level-triggered External trigger mode

ETL

Enable low level-triggered External trigger mode

ETHL

Enable reading captured on a high to low transition

Advertising