User’s Guide


30

If failed:

6. The following screen appears after the offsets have been calculated:

Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press ENTER.

7. The display appears as follows:

Optionally exercise the load cell shaft several times (at full scale, if possible), then press ENTER.

8. The display appears as follows:

Apply a weight equal to the full scale of the instrument, then press ENTER.

9. After displaying “Please wait…” the display appears as follows:

Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.

CALIBRATION

COMPRESSION

Gain adjust

Apply full scale load

10.000 lbF +/-20%,

then press ENTER.

CALIBRATION

COMPRESSION

Ensure no load,

then press ZERO.

CALIBRATION

COMPRESSION

Attach necessary
weight fixtures,
then press ENTER.

CALIBRATION

COMPRESSION

Optionally exercise
sensor, then press
ENTER.

CALIBRATION

OFFSET

Sensor failed

Analog failed

CALIBRATION

OFFSET

Sensor passed
Analog passed