Hudson Robotics SOLO User Manual

Page 21

Advertising
background image

21

b. ‘D’ Syringe Type (Side-mounted valved syringes) –

This option allow the user to move liquid from any of the systems fluid

reservoir vessels to or from the tip. For both Aspirate and Dispense, the

reservoir is selected by selecting the appropriate Valve Port in the

displayed input box:

Aspirate – the user can enter a volume (uL) to move from the tip to the

reservoir (even volumes greater than the syringe’s volume), or by

checking the ‘Continuous’ operation box, run the syringe piston

continuously until the user clicks on the “Halt” button. The ‘No.

Strokes’ box can be checked and the desired number of aspiration strokes

entered as another operation option. If there are multiple syringe pumps

on the SOLO, the Syringe box will drop down to allow the user to select

any individual syringe or, at the bottom of the list, “ALL” syringes

operating together.

Dispense – the user can enter a volume (uL) to move from a reservoir to the

tip (even volumes greater than the syringe’s volume), or by

checking the ‘Continuous’ operation box, run the syringe piston

continuously until the user clicks on the “Halt” button. The ‘No.

Strokes’ box can be checked and the desired number of dispensing

strokes entered as another operation option.