Hudson Robotics SOLO User Manual

Page 54

Advertising
background image

54

The user would enter the desired temperature to be achieved (the SetPoint)

in the appropriate box. The range is -10.0 to 110.0 Deg.C. When a

Method is running, the “Actual:” box will display the actual temperature

of the Nest. The user has several optional ways to execute this step:

Turn Nest OFF at End of Method: un-checking this will allow the

user to keep the heating/cooling action running after the running

Method ends. If this remains checked, it will turn OFF at the end

of a run.

Don’t Wait: this will send the SetPoint to the Nest, starting

heating/cooling action, but then will immediately continue to the

next step in this method.

Wait Until Actual = SetPoint: this will cause this step to hold here

until the Nest’s actual temperature reaches the targeted SetPoint,

then will immediately move on to the next method step.

Wait Until Temperature is Stable: this will cause this step to hold

here until the actual temperature stops changing, even though

SetPoint may have been reached before this.

Turn Nest OFF Now: checking this will simply use this method

step to turn the heating/cooling action OFF, then the step will exit.

Turn Nest OFF at Temperature: this will turn the nest OFF

immediately upon satisfying either of the previous two selections,

then advance to the next method step.

Log Temperature @: checking this box will cause this step to log

the ACTUAL temperature to the SOLOSoft log file at intervals

given by the entry in the accompanying box (in seconds). This

logging will continue even after this step exits, until the method

ends.