Increment Order – (applies only to aspirating from a plate position for

SOLO equipped with single tip pipettor head or multi-tip pipettor

head used in ‘Single Tip Mode’)

Row Order (disabled for multi-tip option of ‘P8’type syringe) : If a

multiple number of wells per execution is entered, or

if this step is within a Loop, each aspiration will increment in the

order “A1, A2, …. , A12, B1,…, B12, etc.”.

Column Order(disabled for multi-tip option of ‘P12’type syringe):

“A1, B1, … , H1, A2, …, H2, etc.”.

Reverse Order: wells with non-zero volumes are selected beginning at

H12 (for a 96-well plate) instead of A1. Row or Column order rules

apply, but in reverse.

Note: For multi-tip ‘P8’ syringe all non-zero volume wells in each

row will be addressed before moving to the next row.

Note: For multi-tip ‘P12’ syringe all non-zero volume wells in each

column will be addressed before moving to the next column.

Any wells with a zero volume will be skipped by single-tip units and

not included in the increment count.

Syringe (uL) –

This is read from the SOLO’s firmware and is selected from a

dropdown list of the syringes actually on this unit.

Start by Emptying Syringe (‘Pn’ syringe type only, where n is 0,1,8 or 12)

Checking this box will cause the needle or tip to move over the

intended aspirate position, then the syringe piston will move to its

bottom (empty) position at the start of this step. This is useful to

ensure that the full capacity of the syringe is available for this

aspiration. Otherwise the aspiration will start at the last position of the

syringe piston.

Find Bottom of Vessel or Well –

If this box is checked, the tip will travel down under low power until it

stalls on the bottom of the plate well or tube. Then, it will rise by the

distance in the Z Shift box under the ‘Aspirate Shift’ parameters.

Mixing –

If the box is checked, immediately after either dispence or aspiration,

the SOLO will dispense then aspirate the amount in ‘Mix Volume’, for

the number of times given by the ‘Mix Cycles’ value, ending with a

final dispense to leave the mixed sample in the well or vessel.

The aspirating action of the mixing will take place at the height

indicated by the ‘Z Shift’ entry, while the dispense action will occur at

the ‘Dispense Height’ distance from the well bottom (for plates)

or from the taught point.

Valve Port (‘D’ syringe type only, not shown) –

The user must select the appropriate fluid reservoir by choosing the

correct valve from a drop down list for the selected syringe pump.