Hudson Robotics SOLO User Manual

Page 36

Advertising
background image

36

3.4 – Get Tip Step

This step, whose GUI is shown in the next figure, will start by shucking any tip

that may have been left on the mandrel from a prior activity, then move to the

next tip position in a tip box located at, in this example, Position1:

The location of the tip box is selected by clicking on the appropriate nest position

in the displayed deck image. At the start of this step, the SOLO will attempt to

shuck the tip (or tips), in case the tips were left on the mandrel from a prior

operation. If the position selected in the “Select Disposal Point” box is the same

as the plate nest where the tip box is located, the tips will be shucked into the last

box position emptied. However, if a method was just loaded, if the tip box is full,

or if a multi-tip unit is being used with fewer than its maximum tip count (i.e., 8

or 12), then the tip shuck will occur at the position in the “Select Disposal Point”

(if it is different from the location of this tip box) or at the “Tip Disposal” position

by default.

For all syringe types SOLOSoft keeps track of tip count in tip boxes between

methods and even if closed and powered-down. For single-tip units the tip

retrieved in this step will be the next tip in the box, counted across the box in ‘row

order’, starting from the H12 (or P24) well position. For multi-tip syringe types

‘P8’ and ‘P12’ SOLOSoft keeps track of tips available in tip box for both single-

tip and multi-tip operations.