ITC Hemochron Signature Elite Whole Blood Microcoagulation System User Manual

Page 2

Advertising
background image

2

INTENDED USE

The HEMOCHRON Signature Elite

®

Whole Blood Microcoagulation System is a battery-operated, hand-held

instrument that performs individual point-of-care coagulation tests on fresh or citrated whole blood. These tests
include: Activated Clotting Time (ACT+ and ACT-LR), Activated Partial Thromboplastin Time (APTT and APTT
Citrate), and Prothrombin Time (PT and PT Citrate). The system is intended to be used only with test cuvettes that
are available from ITC.

Data management capabilities are included with the instrument. These capabilities include storage of up to 600
patient results and 600 quality control results, designation of quality control levels, tagging of test results with
date and time, entry of Patient ID and/or Operator ID or Operator PIN, and printing of results.

HEMOCHRON

®

Configuration Manager software is included. This software allows the user to connect a

personal computer to the instrument and perform system configuration functions using the fast and convenient
Microsoft

®

Windows

®

user interface. HEMOCHRON ReportMaker™ software, provided separately, allows the

user to connect a personal computer to an instrument and perform various data management and data
reporting functions.

For in vitro Diagnostic Use.

SUMMARY AND EXPLANATION

Events that lead to formation of a blood clot are simplified in coagulation theory into two interactive
coagulation cascades that are referred to as the intrinsic and the extrinsic pathways.

The clotting factors involved in these pathways are numbered I through V and VII through XIII. The intrinsic
pathway begins with the contact activation of factor XII and, through the interaction of several coagulation
factors, results in the conversion of soluble fibrinogen to insoluble fibrin strands. The extrinsic pathway is
initiated through the interaction of tissue factor with factor VII. Platelets, essential co-factors in this reaction,
provide the platelet phospholipid surface on which coagulation reactions occur.

Activated Clotting Time (ACT+ and ACT-LR), Activated Partial Thromboplastin Time (APTT and APTT Citrate),
and Prothrombin Time (PT and PT Citrate) tests are general coagulation screening tests that are used to
measure the functionality of the blood coagulation cascade. The ACT test is the method of choice for
monitoring heparin therapy during cardiac surgery and PCI. Fresh whole blood is added to an activator
(Celite

®

, silica, kaolin, or glass particles), and then timed for the formation of a clot. The type of activator used

will affect the degree of prolongation of the ACT to a given heparin dose.

The HEMOCHRON

®

Jr. ACT+ test uses a mixture of silica, kaolin, and phospholipids as an activator to create a

rapid and highly sensitive alternative to existing ACT tests. This test demonstrates linearity at heparin
concentrations ranging from 1.0 to 6.0 units of heparin per mL of blood, and it is not affected by high dose
aprotinin therapy.

The HEMOCHRON Jr. ACT-LR test uses a Celite activator due to its excellent heparin sensitivity. The test
demonstrates linearity at heparin concentrations up to 2.5 units of heparin per mL of blood. The test is not
intended for use with aprotinin therapy.

The HEMOCHRON Jr. APTT test measures the intrinsic coagulation pathway and involves all coagulation factors
except factors VII and III (tissue factor). The APTT test improves the PTT test through use of a contact
activating substance, which standardizes activation of Factor XII to provide a more precise and sensitive assay
for low level heparin monitoring. The HEMOCHRON Jr. APTT test is formulated to provide optimal heparin
sensitivity at heparin concentrations up to 1.5 units of heparin per mL of blood.

The HEMOCHRON Jr. APTT Citrate test performs the same measurement as the APTT test, using a citrated
whole blood sample.

Advertising