ITC Hemochron Signature Elite Whole Blood Microcoagulation System User Manual

Page 41

Advertising
background image

41

b. To print results obtained on a specific date, press 2 to display the Print by Date screen:

Enter the date using the keypad, then press ENTER. All QC results obtained on the specified date
are printed.

c. To print all results for a specific assay, press 3 to display the Print by Assay screen:

Note: If the desired assay is not shown, press 7 to display additional assays.

Enter the number for the assay, then press ENTER. All QC test results for the specified assay are
printed.

d. To print results for a specific cuvette lot number, press 4 to display the Print by CVlot screen:

Scan or type the cuvette lot number, then press ENTER. All QC test results for the specified
cuvette lot number are printed.

e. To print results for a specific control lot number, press 5 to display the Print by QClot screen:

Scan or type the control lot number, then press ENTER. All QC test results for the specified
control lot number are printed.

f. Press CANCEL to return to the menu for printing QC results (Step 1).

To Check Printer Function:

1.

Press 6 to output an ASCII character set to an attached printer to verify that the printer port
functions properly.

HEMOCHRON Elite Print Test
!”#$%&’()*+,-./012345678
9:;<=>?@ABCDEFGHIJKLMNOP
QRSTUVWXYZ[\]^_`abcdefgh
ijklmnopqrstuvwxzy{|}~

To Abort Printing of Results:

1.

Pressing CANCEL for three seconds will abort data transmission to the printer. Any data transmitted
before CANCEL is released will complete printing.

Advertising